Pairing the Blood Pressure Monitor with your Smartphone To begin using Bluetooth® for the rst time, please visit the web-site at http://www.rossmax.com for the initial set-up instructions. 1. Download and install the applicable APP onto your smart-phone.2. To pair this device with your smartphone, turn on the device, Bluetooth® and the App of smartphone, and follow set-up and pairing instructions.3. If the pairing is successful, the Bluetooth® symbol ( ) will appear on the display and keep ashing during data transfer. The current measured value will automatically be transferred to the App when the measurement is completed.4. If the pairing has failed, the Bluetooth® symbol ( ) will not appear on the display and the current measured value will not automatically be transferred to the App. In this case, the value is saved in the selected user memory zone. Please re-pair this device with your smartphone and follow App instructions for Bluetooth® transfer.
APP
APPAPP
APP
120mmHg
120mmHg
USB
PC
BP Manager
BP Manager
1. Adjust the date/time/Bluetooth in the monitor by holding down the ON/OFF/START key for ap-proximately 5 seconds under power o mode. The display will show a blinking number show-ing the year.2. Change the year by pressing the Memory key, each press will increase the number. Press the ON/OFF/START key to conrm the entry and the screen will show a blinking number representing the date.3. Change the date, the hour and the minute as described in Step 2 above, using the Memory key to change and the ON/OFF/START key to conrm the entries.
4. After adjusting the date/time, the Bluetooth® symbol ( ) and the blinking icon “ ” will
be shown on the display simultaneously. Use the Memory key to choose whether automatic Bluetooth® data transfer is activated (Bluetooth® symbol ( )+ ) or deactivated (Bluetooth® symbol ( )+ ) and conrm with the ON/OFF/START key.5. Press the ON/OFF/START key again, “0” will reappear as the Blood Pressure Monitor is ready for measurement.If any abnormality will arise during use, please check the following points. Problems Possible causes SolutionNothing appears on the display when operating the unit.Battery at. Fully charge the battery.Malfunction.Press reset button on the back side of the unit with a paper clip or similar thin object.LED indicator does not illuminate when charging the battery.The AC Adapter has not been cor-rectly inserted into the port of the device or the power socket.Correctly insert the AC Adapter into the port/socket.The monitor or the AC Adapter is abnormally hot when charging the battery.The monitor or the AC adapter may be damaged. Unplug the AC adapter from the socket immediately and contact your local distributor.The unit does not work from its battery.Battery at. Fully charge the battery.Battery ages.If the unit does not operate after fully charging, replace the battery with a new one.Battery broken.If the battery cannot be fully charged and the GREEN charg-ing indicator keeps illuminating, replace the battery with a new one. EE mark shown on display or the blood pressure value is displayed excessively low (high).Is the cu placed correctly? Wrap the cu properly so that it is positioned correctly.Did you talk or move during measurement?Measure again. Keep arm steady during measurement.Did you vigorously shake the cu during measurement.Any abnormality occurs during the measurement.Malfunction.Press reset button on the back side of the unit with a paper clip or similar thin object.Note: If the device is still not working properly, return it to your dealer. Under no circumstance should you disassemble and repair the unit by yourself.
1. The unit contains high-precision assemblies. Therefore, avoid extreme temperatures, hu-
midity, and direct sunlight. Avoid dropping or strongly shocking the main unit, and pro-tect it from dust.2. Clean the blood pressure monitor body and the cu carefully with a slightly damp, soft cloth. Do not press. Do not wash the cu or use chemical cleaner on it. Never use thinner, alcohol or petrol (gasoline) as cleaner.3. The unit should not be operated by children so to avoid hazardous situations.4. If the unit is stored near freezing, allow it to acclimate at room temperature before use.5. This unit is not eld serviceable. You should not use any tool to open the device nor should you attempt to adjust anything inside the device. If you have any problems, please contact the store or the doctor from whom you purchased this unit or please contact Rossmax International Ltd.6. As a common issue for all blood pressure monitors using the oscillometric measurement function, the device may have diculty in determining the proper blood pressure for us-ers diagnosed with diabetes, poor circulation of blood, kidney problems, or for users suf-fered from stroke, or for unconscious users.7. This unit is able to detect common arrhythmia (atrial or ventricular premature beats or atrial brillation). The ARR, AFib and PC icons are displayed after the measurement if Atrial Fibrillation and Premature Contraction was detected during the measurement. If ARR, AFib or PC icons are displayed, you are advised to wait for a while and take another meas-urement. It is strongly recommended that you consult your physician if the ARR, AFib or PC icons appear often.8. While the given device is able to detect specic pulse arrhythmia, the measurement ac-curacy of the blood pressure meter may be impaired with the occurrence of pulse ar-rhythmia. 9. To stop operation at any time, press the ON/OFF/START key, and the air in the cu will be rapidly exhausted.
10. Once the ination reaches 300 mmHg, the unit will start deating rapidly for safety rea-
sons.11. Please note that this unit can be a home healthcare product, but it is not intended to serve as a substitute for the advice of a physician or medical professional.
12. Do not use this device for diagnosis or treatment of any health problem or disease. Meas-
urement results are for reference only. Consult a healthcare professional for interpreta-tion of pressure measurements. Contact your physician if you have or suspect any medi-cal problem. Do not change your medications without the advice of your physician or healthcare professional.13. Electromagnetic interference: The device contains sensitive electronic components. Avoid strong electrical or electromagnetic elds in the direct vicinity of the device (e.g. mobile telephones, microwave ovens) or less than 1.5 km from AM, FM or TV broadcast antennas. These may lead to temporary impairment of measurement accuracy.14. Dispose of device, batteries, components and accessories according to local regulations.15. This monitor may not meet its performance specication if stored or used outside tem-perature and humidity ranges specied in Specications.16. Please note that when inating, the functions of the limb in question may be impaired.17. During the blood pressure measurement, blood circulation must not be stopped for an unnecessarily long time. If the device malfunctions, remove the cu from the arm.18. Avoid any mechanical restriction, compression or bending of the cu line.19. Do not allow sustained pressure in the cu or frequent measurements. The resulting re-striction of the blood ow may cause injury.20. Ensure that the cu is not placed on an arm in which the arteries or veins are undergoing medical treatment, e.g. intravascular access or therapy, or an arteriovenous (AV) shunt.21. Do not apply the cu on the side, where a mastectomy has been performed in your patient history.22. Do not place the cu over wounds as this may cause further injury.23. Only ever use the cus provided with the monitor or original replacement cus. Other-wise erroneous results will be recorded.24. Batteries can be fatal if swallowed. You should therefore store the batteries and products where they are inaccessible to small children. If a battery has been swallowed, call a doc-tor immediately.25. Do not use the tubing and/or AC Adapter for any other purpose than those specied, as they can cause risk of strangulation.26. Do not service or maintain device and cu while in use. 27. This unit should not be used adjacent to or stacked with other equipment.28. Please do not use any other cables or accessories not approved by the manufacturer in this manual to avoid negative inuence on electromagnetic compatibility. Measurement Method OscillometricMeasurement Range Pressure: 30~260 mmHg; Pulse: 40~199 beats/ minutePressure SensorSemi conductor
Accuracy Pressure: ± 3mmHg; Pulse: ± 5% of reading
InflationPump DrivenDeflationAutomatic Air Release ValveMemory capacity60 memories for each zone x 2 zonesAuto-shut-off1 minute after last key operationPermissible Operating Temperature and Humidity10°C~40°C (50°F~104°F); 15%~85% RH; 700~1060
hPa
Permissible Transport and Storage Temperature and Humidity-10°C~50°C (14°F~122°F); 10%~90% RH; 700~1060
hPa
Adapter Input100-240V, 50/60 HzAdapter Output / USB Input5V 1A (Type C)Li-lon BatteryDC 3.7V Dimensions120 (L) X 80 (W) X 57 (H) mmWeight276.0g (with battery, w/o cuff)Arm circumference Adult: 24~40 cm (9.4”~15.7”)Limited Users Adult usersType BF: Device and cuff are designed to provide special protection against electrical shocks.IP ClassificationIP21: Protection against harmful ingress of water and particulate matter* Specifications are subject to change without notice.Time and Bluetooth® AdjustmentData Transfer via Bluetooth®TroubleshootingCautionary NotesSpecications
APP
APP
APP
APP
120mmHg
120mmHg
USB
PC
BP Manager
BP Manager
Electromagnetic Compatibility Information
1. This device needs to be installed and put into service in accordance with the information provided
in the user manual.
2. WARNING: Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part of the Z5, including cables specied by the manufacturer. Otherwise, degradation of the performance of this device could result.If higher IMMUNITY TEST LEVELS than those specied in Table 9 are used, the minimum separation distance may be lowered. Lower minimum separation distances shall be calculated using the equation specied in 8.10.Manufacturer’s declaration-electromagnetic immunityThe Z5 is intended for use in the electromagnetic environment specied below. The customer or the user of the Z5 should assure that is used in such and environment.Immunity
test
IEC 60601 test levelCompliance levelElectromagnetic environment-guidance
Conducted RF IEC 61000-4-63 Vrms: 0,15 MHz – 80 MHz6 Vrms: in ISM and amateur radio bands between 0,15 MHz and 80 MHz 80 % AM at 1 kHz 3 Vrms: 0,15 MHz – 80 MHz6 Vrms: in ISM and amateur radio bands between 0,15 MHz and 80 MHz 80 % AM at 1 kHz Portable and mobile RF communications equipment should be used no closer to any part of the Z5 including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.Recommended separation distance:d = 1,2 √P, d = 1,2 √P 80MHz to 800 MHz, d = 2,3 √P 800MHz to 2,7 GHz Where P is the maximum output power rating of the transmitter in watts (W) ac-cording to the transmitter manufacturer and d is the recommended separation distance in metres (m).Interference may occur in the vicinity of equipment marked with the following symbol: Radiated RF IEC 61000-
4-3
10 V/m80 MHz – 2,7 GHz 80 % AM at 1 kHz10 V/m80 MHz – 2,7 GHz 80 % AM at 1 kHzNOTE1: At 80 MHz and 800 MHz, the higher frequency range applies.NOTE2: These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reection from structures, objects and people.More information on EMC compliance of the device can be obtained from Rossmax website: www.rossmax.com.Notes: 1. Unpairing your device will not delete the data from the App. 2. If you re-pair your smartphone with your blood pressure monitor, all prior reading history stored on the App will be retained.3. Bluetooth® data transfer will reduce the battery capacity.1. Press the User-Switching key to select memory zone 1 or memory zone 2.2. Press and hold the Memory key for approximately 5 seconds, then the data in the memory zone can be erased automatically.Clearing Values from MemoryIN0Z50000000000XXOBM_IB_Z5(1)NA19BT_EN_SWCMC_ver2337
WARNING: The symbol on this product means that it's an electronic product and
following the European directive 2012/19/EU the electronic products have to be dis-posed on your local recycling centre for safe treatment.
Model: Z5
www.rossmax.com
Warranty CardThis instrument is covered by a 5 year guarantee from the date of purchase. The guar-antee is valid only on presentation of the warranty card completed or stamped by the seller/dealer confirming date of purchase or the receipt. Batteries, cuff and accessories are not included. Opening or altering the instrument invalidates the guarantee. The guarantee does not cover damage, accidents or non-compliance with the instruction manual. Please contact your local seller/dealer or www.rossmax.com.Customer Name: ____________________________________________________Address: ___________________________________________________________Telephone: __________________________________________________________E-mail address: ______________________________________________________Product InformationDate of purchase: ___________________________________________________Store where purchased: ____________________________________________________________________
Blood Pressure Monitor
- The Bluetooth® word mark and logos are registered trademarks owned by the Bluetooth SIG, Inc. and any use of such marks by Rossmax International Ltd. is under
license. Other trademarks and trade names are those of their respective owners.
- The blood pressure monitor uses Bluetooth®(Bluetooth® low energy technology)
- Apple and the Apple logo are trademarks of Apple Inc., registered in the U.S. and other countries. App Store is a service mark of Apple Inc.
- Google Play and the Google Play logo are trademarks of Google Inc.Data Transfer via Bluetooth®Please download and install the free Healthstyle APP onto your smartphoneHealthstyle APPRossmax Swiss GmbH Widnauerstrasse 1, CH-9435 Heerbrugg, SwitzerlandCMC Medical Devices & Drugs S.L.C/Horacio Lengo N° 18, CP 29006, Málaga, Spain
View the manual for the Rossmax Z5 here, for free. This manual comes under the category sphygmomanometer and has been rated by 1 people with an average of a 8.6. This manual is available in the following languages: English. Do you have a question about the Rossmax Z5 or do you need help? Ask your question here
Do you have a question about the Rossmax and is the answer not in the manual?
Receive updates on solutions
Get answers to your question
Question and answer
Number of questions: 0
The Rossmax Z5 sphygmomanometer is a medical device used for measuring blood pressure. It features a clear display screen to show accurate readings of both systolic and diastolic pressure, as well as the pulse rate. The device is equipped with a comfortable cuff that can fit a wide range of arm sizes, ensuring accurate results for a variety of patients. The Rossmax Z5 is designed with precision and accuracy in mind, making it a reliable tool for healthcare professionals. The device is portable and easy to use, allowing for convenient blood pressure monitoring in a clinical setting or at home. Overall, the Rossmax Z5 sphygmomanometer is a dependable and user-friendly device for accurately measuring blood pressure.
Can't find the answer to your question in the manual? You may find the answer to your question in the FAQs about the Rossmax Z5 below.
What does systolic pressure mean?
Systolic pressure is the top number that the blood pressure monitor will indicate. Systolic pressure refers to the amount of pressure in your arteries during the contraction of your heart muscle.
What does diastolilc pressure mean?
Diastolic pressure is the lowest number that the blood pressure monitor will indicate. The bottom number refers to your blood pressure when your heart muscle is between beats.
What is a normal blood pressure?
A normal blood pressure is about 120/80.
Is the manual of the Rossmax Z5 available in English?
Yes, the manual of the Rossmax Z5 is available in English .
Need help?
Do you have a question about the Rossmax and is the answer not in the manual?
Question and answer
Number of questions: 0