Rossmax SD100 manual

Rossmax SD100
9.4 · 1
PDF manual
 · 2 pages
English
manualRossmax SD100

Cleaning

Maintenance

Troubleshooting

Specification

1. Please clean the surface of the device before using. Wipe the device with medical alcohol (70% (w/w) Ethanol) first, and then let it dry in air or clean it by dry clean fabric. When cleaning the device with water, the water temperature should be lower than
60°C
2. Using the medical alcohol to disinfect the product after use, prevent from cross infection for next time
use.
3. The best storage environment of the device is - 25°C to 70°C ambient temperature and not higher than 90% relative humidity. Note: 1. Do not sterilize, autoclave or immerse this de-vice in liquid. Do not pour or spray any liquids onto the device. 2. Do not use caustic or abrasive cleaning agents, or any cleaning agent containing am-monium chloride or isopropyl alcohol.Recommends user to return this device to the man-ufacturer perform the following checks every 24 months.- Inspect the equipment for mechanical and function-al damage or deterioration.- Ensure all user interface keys and accessories func-tion normally.Note: Manufacturer use Index2 SpO2 simulator to verify operation of the pulse oximeter device.Symptoms Check points CorrectionsSpO2 or pulse rate cannot displayedApplied nger improperly.Place the nger prop-erly and try again.SpO2 is too low to detectTry again; go to con-sult with your physi-cian if you are sure the device works well.SpO2 or pulse rate are not displayed stablyApplied nger improperly.Place the nger prop-erly and try again.Finger is shaking or body is moving.keep body steadyNo display when button is pressedBatteries run
down
Replace with new batteriesBatteries not inserted correctly.Re-insert batteriesThe display disappears suddenlyThe device will auto power o when it gets no signal.NormalLow batteryReplace with new batteries Note: If the unit does not work, return it to your dealer. Under no circumstance should you disassemble and repair the unit by yourself.SpO2 Measuring range35%~99%, (the resolution is 1%).Accuracy 70%~99%: ±2%, Below 35~69%: unspecied.Optical Sensor The wavelength of red LED is 660 nm and Infrared LED is 905/880 nm with maximum optical output power of 4 mW/sr.Pulse Measuring range30 bpm~250 bpm (the resolution is 1
bpm)
Accuracy ±3 bpmPower source AAA × 2 (Alkaline)Battery life Continually for 16 hours with two alkaline batteriesOperating ConditionTemperature: 5°C~35°C (41°F ~ 95°F), Relative Humidity: 15-95% (non condensing), Atmospheric pressure: 700hPa ~ 1060hPa, Attitude: -1,280 to 12,000 feet (-390m to 3,658m)Storage / Transportation ConditionTemperature: -25°C~+70°C(-13°F ~ 158°F), Relative humidity: 15-90%(non condensing), Atmospheric pressure: 700hPa ~ 1060hPa, Attitude: -1,280 to 12,000 feet (-390m to 3,658m), The time from 70°C or -25°C back to use : 3 hoursDimensions 61.3(L) × 31W) × 36.7(H) mmWeight About 33g (without the batteries)Standards IEC60601-1-2, Class B, IEC60601-1, Type BF, ISO80601-2-61, IEC/EN60601-1-11Type BF applied partsIP Classication IP22: Protection against harmful ingress of water and particulate matterNote: - A description of the effect on displayed and transmit-ted SpO2 and pulse rate:- Data averaging: 4 seconds for SpO2; 8 seconds for pulse rate.- Data update delay: Less than 2 seconds.be evaluated at the measurement site. Rather, they are exclusively used to display the current signal vari-ation at the measurement site and do not enable re-liable diagnostics for the pulse.- A warning that other cables and accessories may negatively aect EMC performance.- A warning regarding stacking and location close to other equipment.- A warning that use of other accessories results in non-compliance.- The safety way in high-temperature environment for all people use is measuring for 10 minutes, and turn it o for 20 minutes before measure again.- The oximeter is calibrated in the factory before sale, there is no need to calibrate it during its life cycle.
Model: SD100

www.rossmax.com

Fingertip Pulse Oximeter
Rossmax InnoTek Corp. 12F., No. 189, Kang Chien Rd., Taipei, 114, Taiwan.CMC Medical Devices & Drugs S.L.C/ Horacio Lengo Nº 18, CP 29006, Málaga, Spain
WARNING: The symbol on this product means that it's an electronic product
and following the European directive 2012/19/EU the electronic products have to be dispose on your local recycling centre for safe treatment.EMC guidance and manufacturer’s declaration
Recommended separation distances between portable and mobile RF communications equipment and the ME equipment
The Finger-tip pulse oximeter is intended for use in an electromagnetic environment in which radiated RF disturbances are con-
trolled. The customer or the user of the Finger-tip pulse oximeter can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and the Finger-tip pulse oximeter
as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter / W
Separation distance according to frequency of transmitter / m
150 kHz to 80 MHz , d=[3.5/
V1]√P
80 MHz to 800 MHz , d=[3.5/
E1]√P
800 MHz to 2,5 GHz , d=[7/
E1]√P
0.01 0.1 0.1 0.2
0.1 0.4 0.4 0.7
1 1.2 1.2 2.3
10 3.7 3.7 7.4
100 11.7 11.7 23.3
Declaration – electromagnetic emissions
The Finger-tip pulse oximeter is intended for use in the electromagnetic environment specied below. The customer or the user of
the Finger-tip pulse oximeter should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions CISPR 11 Group 1 Portable and mobile RF communications equipment should be used no
closer to any part of the EQUIPMENT or SYSTEM including cables, than the
recommended separation distance calculated from the equation applicable
to the frequency of the transmitter. Interference may occur in the vicinity of
equipment marked with the following symbol.
RF emissions CISPR 11 Class B The Finger-tip pulse oximeter is suitable for use in all establishments,
including domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings used
for domestic purposes.
Harmonic emissions IEC 61000-3-2 N/A
Voltage uctuations/Flicker emissions
IEC 61000-3-3
N/A
Declaration – electromagnetic emissions and immunity – for EQUIPMENT and SYSTEMS that are use in the professional health-
care facility environment or in the home healthcare environment
The Finger-tip pulse oximeter declaration – electromagnetic immunity
The Finger-tip pulse oximeter system is intended for use in the electromagnetic environment specied below. The customer or
the user of the Finger-tip pulse oximeter system should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Conducted RF IEC
61000-4-6
3 Vrms ; 6 Vrms ; 150
kHz to 80 MHz
N/A Portable and mobile RF communications equipment
should be used no closer to any part of the EQUIPMENT
or SYSTEM including cables, than the recommended
separation distance calculated from the equation ap-
plicable to the frequency of the transmitter. Interference
may occur in the vicinity of equipment marked with the
following symbol.
Radiated RF IEC
61000-4-3
3 V/m ; 10V/m ; 80
MHz – 2.7 GHz l 80%
3 V/m ; 10V/m ; 80
MHz – 2.7 GHz ; 80%
Proximity elds
from RF wireless
Communications
equipment IEC
61000-4-3
27 V/m 385 MHz 27 V/m 385 MHz
28 V/m 450 MHz 28 V/m 450 MHz
9 V/m 710 MHz 9 V/m 710 MHz
745 MHz 745 MHz
780 MHz 780 MHz
28 V/m 810 MHz 28 V/m 810 MHz
870 MHz 870 MHz
930 MHz 930 MHz
28 V/m 1720 MHz 28 V/m 1720 MHz
1845 MHz 1845 MHz
1970 MHz 1970 MHz
28 V/m 2450 MHz 28 V/m 2450 MHz
9 V/m 5240 MHz 9 V/m 5240 MHz
5500 MHz 5500 MHz
5785 MHz 5785 MHz
Declaration – electromagnetic immunity
The Finger-tip pulse oximeter system is intended for use in the electromagnetic environment specied below. The customer or
the user of the Finger-tip pulse oximeter system should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV,
±15 kV air
±8 kV contact
±2 kV, ±4 kV, ±8
kV, ±15 kV air
Floors should be wood, concrete or ceramic tile.
If oors are covered with synthetic material, the
relative humidity should be at least 30 %.
Electrical fast transient/
burst IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
N/A Mains power quality should be that of a typical
commercial or hospital environment.
Surge IEC 61000-4-5 ±0.5 kV
±1 kV dierential mode
±2 kV common mode
N/A Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power
supply input lines IEC
61000-4-11
0 % U
T
; 0, 5 cycle At 0°, 45°,
90°, 135°, 180°, 225°, 270°
and 315°
0 % U
T
; 1 cycle And 70
% U
T
; 25/30 cycle Single
phase: at 0°
N/A Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the EQUIPMENT or SYSTEM requires continued
operation during power mains interruptions, it is
recommended that the EQUIPMENT or SYSTEM
be powered from an uninterruptible power sup-
ply or a battery.
Power frequency (50/60
Hz) magnetic eld IEC
61000-4-8
30 A/m 30 A/m Power frequency magnetic elds should be at
levels characteristic of a typical location in a typi-
cal commercial or hospital environment.
Warranty CardThis instrument is covered by a 2 years guarantee from the date of purchase, batteries and accessories are not included. The guarantee is valid only on presentation of the guarantee card completed by the dealer confirming date of purchase or the receipt. Opening or altering the instrument invalidates the guarantee. The guarantee does not cover damage, accidents or non-compliance with the instruction manual. Please contact your local seller/dealer or www.rossmax.com.Customer Name: _______________________________Address: ____________________________________Telephone: __________________________________E-mail address: ________________________________Product Information: ___________________________Date of purchase: ______________________________Store where purchased: ___________________________________________IN0SD100000000011RI_IB_SD100_EN_TP_ver2214

View the manual for the Rossmax SD100 here, for free. This manual comes under the category heart rate monitor and has been rated by 1 people with an average of a 9.4. This manual is available in the following languages: English. Do you have a question about the Rossmax SD100 or do you need help? Ask your question here

Need help?

Do you have a question about the Rossmax and is the answer not in the manual?

  • Receive updates on solutions
  • Get answers to your question

Question and answer

Number of questions: 0

The Rossmax SD100 heart rate monitor is a medical device used to measure an individual's heart rate. It is designed to be user-friendly and accurate, providing reliable readings for those in need of monitoring their heart rate on a regular basis. The device is equipped with a clear display screen that allows for easy reading of the heart rate measurements. Additionally, the Rossmax SD100 heart rate monitor is compact and lightweight, making it convenient to carry around and use wherever and whenever needed. The device also has a long battery life, ensuring that it can be used for extended periods of time without needing frequent recharging. Overall, the Rossmax SD100 heart rate monitor is a reliable and efficient tool for individuals looking to monitor their heart rate for medical or fitness purposes. With its user-friendly design and accurate measurements, it is a valuable device for those in need of regular heart rate monitoring.

General
BrandRossmax
ModelSD100
Productheart rate monitor
LanguageEnglish
FiletypeUser manual (PDF), Brochure (PDF)

Can't find the answer to your question in the manual? You may find the answer to your question in the FAQs about the Rossmax SD100 below.

How much should I exercise as an adult per week?

As an adult, it is recommended to do moderately intensive exercise for at least 2.5 hours per week. Preferably spread over several days.

How high is an adult's heart rate at rest?

An adult's resting heart rate is between 60 and 100 beats per minute.

What is an adult's maximum heart rate?

As a rule of thumb for your maximum heart rate, you can use 220 minus your age.

Is the manual of the Rossmax SD100 available in English?

Yes, the manual of the Rossmax SD100 is available in English .

Is your question not listed? Ask your question here

No results