Rossmax SB220 manual

Rossmax SB220
8.2 · 1
PDF manual
 · 2 pages
English
manualRossmax SB220

Cleaning

Maintenance

Troubleshooting

Specification

1. Please clean the surface of the device before using. Wipe the device with medical alcohol (70% isopro-pyl alcohol)first, and then let it dry in air or clean it by dry clean fabric. When cleaning the device with water, the water temperature should be lower than
60°C
2. Using the medical alcohol to disinfect the product after use, prevent from cross infection for next time
use.
3. The best storage environment of the device is - 25°C to 70°C ambient temperature and not higher than 90% relative humidity. Note: 1. Do not sterilize, autoclave or immerse this device in liquid. Do not pour or spray any liquids onto the device.2. Do not use caustic or abrasive cleaning agents, or any cleaning agent containing ammonium chloride or isopropyl alcohol.Recommends user to return this device to the man-ufacturer perform the following checks every 24 months.- Inspect the equipment for mechanical and functional damage or deterioration.- Ensure all user interface keys and accessories function normally.Note: Manufacturer use Index2 SpO2 simulator to ver-ify operation of the pulse oximeter equipment.Symptoms Check points CorrectionsSpO2 or pulse rate cannot displayedApplied nger improperly.Place the nger prop-erly and try again.SpO2 is too low to detectTry again; go to consult with your physician if you are sure the device works well.SpO2 or pulse rate are not dis-played stablyApplied nger improperly.Place the nger prop-erly and try again.Finger is shaking or body is moving.Keep body steadyNo display when button is pressedBatteries run downReplace with new batteriesBatteries not inserted correctly.Re-insert batteriesThe display disappears suddenlyThe device will auto power o when it gets no signal.NormalLow batteryReplace with new batteries
SpO
2
Measuring range 35%~99%, (the resolution is 1%)Accuracy 70%~99%: ±2%, Below 35~69%: unspeciedOptical Sensor The wavelength of red LED is 660 nm and Infrared LED is 905/880 nm with maximum optical output power of 4 mW/sr.Pulse Measuring range 30 bpm~250 bpm (the resolution is 1 bpm)Accuracy ±3 bpmPower source AAA × 2 (Alkaline)Battery life Continually for 16 hours with two alkaline batteriesOperating ConditionTemperature: 5°C~40°C (41°F ~ 104°F), Relative Humidity: 15-95% (non condensing), Atmospheric pressure: 700hPa ~ 1060hPa, Attitude: -1,280 to 12,000 feet (-390m to 3,658m)Storage / Transportation ConditionTemperature: -25°C~+70°C(-13°F ~ 158°F), Relative humidity: 15-90%(non condensing), Atmospheric pressure: 700hPa ~ 1060hPa, Attitude: -1,280 to 12,000 feet (-390m to 3,658m)Dimensions 63.5(L) × 34(W) × 35(H) mmWeight About 37g (without the batteries)Standards IEC60601-1-2, Class B, IEC60601-1, Type BF, ISO80601-2-61, IEC60601-1-11Type BF applied partsIP Classication IP22: Protection against harmful ingress of water and particulate matterAttentionSerial numberManufacturerDate of manufactureAuthorized representative in the European CommunityNote:- A description of the effect on displayed and transmitted SpO2 and pulse rate:- Data averaging: 4 seconds for SpO2; 8 seconds for pulse rate.- Data update delay: Less than 2 seconds.- The oximeter is calibrated in the factory before sale. There is no need to calibrate during its life cycle. Note: If the unit does not work, return it to your dealer. Under no circumstance should you disassemble and repair the unit by yourself.
Model: SB220

www.rossmax.com

Fingertip Pulse Oximeter
Rossmax InnoTek Corp.
12F., No. 189, Kang Chien Rd., Taipei,
114, Taiwan.
CMC Medical Devices & Drugs S.L.
C/ Horacio Lengo Nº 18, CP 29006,
Málaga, Spain
WARNING: The symbol on this product means that it's an electronic product
and following the European directive 2012/19/EU the electronic products have to be dispose on your local recycling centre for safe treatment.
IN0SB221000000013
Warranty CardThis instrument is covered by a 2 year guarantee from the date of purchase, batteries and accessories are not included. The guarantee is valid only on presentation of the guarantee card completed by the dealer confirming date of purchase or the receipt. Opening or altering the instrument invalidates the guarantee. The guarantee does not cover damage, accidents or non-compliance with the instruction manual. Please contact your local seller/dealer or www.rossmax.com.Customer Name: __________________________________Address: _______________________________________Telephone: _____________________________________E-mail address: ___________________________________Product Information: ______________________________Date of purchase: _________________________________Store where purchased: ______________________________________________
RI_IB_SB220_EN_
2Yr_TP_ver2128
Note: 1. The product is with a service life of 5 years in the course of regular use.2. If any serious incident (e.g. death) has occurred in relation to the device should be reported to the dealer, manufacturer, and the competent authority of the Member State in which the user and/or patient is established.3. The text is subject to change without further notice.EMC guidance and manufacturer’s declaration
Recommended separation distances between portable and mobile RF communications equipment and
the ME equipment
The Finger-tip pulse oximeter is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Finger-tip pulse oximeter can help prevent elec-
tromagnetic interference by maintaining a minimum distance between portable and mobile RF communi-
cations equipment (transmitters) and the Finger-tip pulse oximeter as recommended below, according to
the maximum output power of the communications equipment.
Rated maximum output
power of transmitter / W
Separation distance according to frequency of transmitter / m
150 kHz to 80 MHz ,
d=[3.5/V1]√P
80 MHz to 800 MHz ,
d=[3.5/E1]√P
800 MHz to 2,5 GHz ,
d=[3.5/E1]√P
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
1 1.17 1.17 2.33
10 3.7 3.7 7.37
100 11.67 11.67 23.33
Declaration – electromagnetic emissions and immunity – for EQUIPMENT and SYSTEMS that are not LIFE-
SUPPORTING and are specied for use only in a shielded location
The Finger-tip pulse oximeter declaration – electromagnetic immunity
The Finger-tip pulse oximeter system is intended for use in the electromagnetic environment specied
below. The customer or the user of the Finger-tip pulse oximeter system should assure that it is used in
such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment-guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3V
Portable and mobile RF communications
equipment should be used no closer to any part
of the EQUIPMENT or SYSTEM including cables,
than the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter. Interference may
occur in the vicinity of equipment marked with
the following symbol.
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3V/m
Declaration – electromagnetic immunity
The Finger-tip pulse oximeter system is intended for use in the electromagnetic environment specied
below. The customer or the user of the Finger-tip pulse oximeter system should assure that it is used in
such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment-
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete
or ceramic tile. If oors are covered
with synthetic material, the
relative humidity should be at
least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power sup-
ply lines
±1 kV for input/
output lines
±2 kV for power sup-
ply lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge IEC
61000-4-5
± 1kV dierential mode
± 2kV common mode
± 1kV dierential mode
± 2kV common mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips,
short interrup-
tions and voltage
variations on
power supply
input lines IEC
61000-4-11
<5% UT(>95% dip in
UT) for 0,5 cycle
40% UT(60% dip in UT)
for 5 cycles
70% UT(30% dip in UT)
for 25 cycles
<5% UT(>95% dip in
UT) for 5 s
<5% UT(>95% dip in
UT) for 0,5 cycle
40% UT(60% dip in UT)
for 5 cycles
70% UT(30% dip in UT)
for 25 cycles
<5% UT(>95% dip in
UT) for 5 s
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the EQUIPMENT or SYSTEM
requires continued operation
during power mains interruptions,
it is recommended that the EQUIP-
MENT or SYSTEM be powered
from an uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz)
magnetic eld
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic elds
should be at levels characteristic of
a typical location in a typical com-
mercial or hospital environment.
Declaration – electromagnetic emissions
The Finger-tip pulse oximeter is intended for use in the electromagnetic environment specied below. The
customer or the user of the Finger-tip pulse oximeter should assure that it is used in such an environment.
Emission test Compliance Electromagnetic environment-guidance
CE emissions CISPR11 Group 1 The Finger-tip pulse oximeter uses RF energy only for
its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference
in nearby electronic equipment.
RF emissions CISPR 11 Class B The Finger-tip pulse oximeter is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions IEC 61000-3-2 Class A
Voltage uctuations/icker emis-
sions IEC 61000-3-3
Complies

View the manual for the Rossmax SB220 here, for free. This manual comes under the category heart rate monitor and has been rated by 1 people with an average of a 8.2. This manual is available in the following languages: English. Do you have a question about the Rossmax SB220 or do you need help? Ask your question here

Need help?

Do you have a question about the Rossmax and is the answer not in the manual?

  • Receive updates on solutions
  • Get answers to your question

Question and answer

Number of questions: 0

The Rossmax SB220 is a heart rate monitor that accurately measures and displays the user's heart rate in real-time. It is equipped with a reliable sensor that provides precise readings, allowing individuals to monitor their heart health effectively. The monitor is designed with user-friendly features, such as a clear display screen and easy-to-use buttons, making it simple for users to operate. In addition to displaying heart rate measurements, the Rossmax SB220 also includes features such as a stopwatch function and a calorie counter, providing users with comprehensive fitness tracking capabilities. The monitor is lightweight and portable, making it convenient for users to carry and use during physical activities. Overall, the Rossmax SB220 is a practical and functional heart rate monitor that helps individuals track their heart rate and monitor their fitness levels with ease.

General
BrandRossmax
ModelSB220
Productheart rate monitor
LanguageEnglish
FiletypeUser manual (PDF), Brochure (PDF)

Can't find the answer to your question in the manual? You may find the answer to your question in the FAQs about the Rossmax SB220 below.

How much should I exercise as an adult per week?

As an adult, it is recommended to do moderately intensive exercise for at least 2.5 hours per week. Preferably spread over several days.

How high is an adult's heart rate at rest?

An adult's resting heart rate is between 60 and 100 beats per minute.

What is an adult's maximum heart rate?

As a rule of thumb for your maximum heart rate, you can use 220 minus your age.

Is the manual of the Rossmax SB220 available in English?

Yes, the manual of the Rossmax SB220 is available in English .

Is your question not listed? Ask your question here

No results