Measuring range 35% – 100% (the resolution is 1%)Accuracy 70% - 100%: ±2 %, 35% - 69%: unspeciedPulse Rate Measuring range 30 – 250 bpm (the resolution is 1 bpm)Accuracy 30 – 250 ± 3 digitsProbe TypeProbe model Rossmax PA100, PB100, PC100, PD100(Single Use), PF100 Extension cord Rossmax PE100Optical Sensor The wavelength of red LED is 660nm and Infrared LED is 905/880nm with maximum optical output power of 4mW.Electrical SpecicationBattery AA * 4 (Alkaline)Battery Life Continually for 15 hours with 4 alkaline batteriesAC Adaptor Model: HK-X205-A06, HK-XW05-A06, (W=1,2,3,4), HKKS-13116, HKKS-13117Input: AC100-240V, 50/60Hz, 0.2A max; Output: DC 6V, 0.8AEnvironmental conditionsOperation Condition Temperature: 5°C – 40°C (41°F – 104°F), Relative Humidity:15% – 95%(non condensing), Atmospheric pressure: 700hPa ~ 1060hPaStorage /Transport ConditionTemperature: -25°C – 70°C (-13°F – 158°F), Relative Humidity:15% – 90%(non condensing), Atmospheric pressure: 700hPa ~ 1060hPaNote: The condition of -25°C or 70°C back to use should stand for 3 hours at room temperature.Dimension Size: 14.5(L) x 7.25(W) x 2.25cm(H)Weight About 150g (without the batteries)StandardIEC/EN60601-1, IEC/EN60601-1-2, IEC/EN60601-1-11, ISO80601-2-61Symbol DescriptorsManufacturerSerial numberEU representativeType BF (Body Floating)IP ClassicationIP22: Protected against foreign objects and moistureCE MarkWarning: the symbol on this product means that it’s an electronic product and following the European directive 2012/19/EU the electronic products have to be dispose on your local recycling centre for safe treatment. SymptomsCheck pointsCorrectionsSpO2 or Pulse rate cannot displayed
-
-
- -
% SpO
2
Artery
Check
BPM
ACTPPG
The icon “- -“ shows on the screenPlace the nger properly and try again.
ACTPPG
This icon means probe connect failure.Be sure ”Rossmax” probe is connected to the device correctly.
ACTPPG
This icon means probe dysfunctionReplace with new probe.Applied nger improperlyPlace the nger properly and try againSpO2 or Pulse rate are not displayed stablyFinger is shaking or body is moving Keep body steadyApplied nger improperlyPlace the nger properly and try againNo display when the
bottom is pressedBatteries run downReplace with new batteriesBatteries inserted incorrectlyRe-insert batteries Note: After clearing data, the icon will show 0 data.Caution: The deleted data could not be restored.WarningCleaningMaintenanceMemories mode- Save CycleMemories mode- Clear DataSpecification Troubleshooting Audio Signals • This device is not intended for use by people (including children) with restricted physical, sensory or mental skills or a lack of experience and/or a lack of knowledge, unless they are supervised by a per- son who has responsibility for their safety or they receive instructions from this person on how to use the device. Children should be supervised around the device to ensure they do not play with it.• This device only for spot-checking, but not medical result evaluation.
• This device is designed to determine the percentage of arterial oxygen saturation of functional
hemoglobin. Factors that may degrade pulse oximeter performance or affect the accuracy of the measurement include the following:
- Do not apply the pulse oximeter on the same arm as a blood pressure cuff, arterial catheter or
infusion line (s) - Excessive light, such as sunlight or direct home lighting. - Not steady at the site of application (e.g. term-bling) - Moisture in the device - Improperly applied device - Finger is too large or too small to fit into the device - Poor pulse quality - Venous pulsations - Anemia or low hemoglobin concentrations - Cardio green and other intravascular dyes - Carboxyhemoglobin - Methemoglobin - Dysfunctional hemoglobin - Artificial nail or fingernail polish - On fingers with anatomical changes, oedemas, scars or burns.
- The conditional of probe. Use only the Rossmax approved pulse oximeter sensor, cable and
accessories. These parts are not reprocessed. Use of other sensors, cable and accessories can result in inaccurate readings.• Using the device for long periods may cause pain for people with circulatory disorders. Reposition the device (probe) at least once every 4 hours to allow the patient’s skin to breath and to check patient’s condition regularly. • Do not use the device near flammable or explosive gas mixtures.• Do not use the device during an MRI or CT scan.• The device will be affected by electromagnetic interference during operation.• A warning that other cables and accessories may negatively affect EMC performance.
• The device may not work when circulation is reduced. Warm or rub the finger, or re-position the
device.
• This device is a precision electronic instrument and must be repaired by qualified technical
professionals. Field repair of the device is not possible. Do not attempt to open the case or repair the electronics. Opening the case may damage the device and void the warranty.• Do not overextend the device’s spring.• A functional tester cannot be used to access the accuracy of a pulse oximeter monitor.
• Do not self-diagonse or self-medicate on the basis of the measurements without consulting your
doctor. In particular, do not start taking any new medication or change the type and/or dosage of any existing medication without prior approval.
• Do not look directly inside the housing during the measurement. The red light and the invisible
infra-red light in the probe are harmful to your eyes. • Please be aware that user with susceptible skin.
• As with all medical equipment, carefully route patient cabling to reduce the possibility of patient
entanglement or strangulation.• The maximum temperature might reach 42.4°C when operating for long time.
• The oximeter is calibrated in the factory before sale, there is no need to calibrate it during its life
cycle.• Do not reuse or disinfect disposable SpO2 probe. (only for PD100)
1. Please clean the surface of the device before using. Wipe the device with medical alcohol (70%
isopropyl alcohol) first, and then let it dry in air or clean it by dry clean fabric.
2. Using the medical alcohol to clean the product after use, prevent from cross infection for next
time use.
3. The best storage environment of the device is -25°C ~ 70°C ambient temperature and not higher
than 90% relative humidity.
Note: 1. Do not sterilize, autoclave or immerse this device in liquid. Do not pour or spray any liquids
onto the device.
2. Do not use caustic or abrasive cleaning agents, or any cleaning agent containing ammonium
chloride or isopropyl alcohol. Recommends user to return this device to the manufacturer perform the following checks every 24 months.• Inspect the equipment for mechanical and functional damage or deterioration.• Ensure all user interface keys and accessories function normally.Note: Manufacturer use Index2 SpO2 simulator to verify operation of the pulse oximeter device. Optional of memory interval: 60 / 30 / 10 seconds• Press the
button and / button to highlight “Memory” option, and then press the
button to select the option.• Entry “Memory Mode”, press the / button to highlight “Save Cycle” option, and then press
the
button to select the option.• Entry “Save Cycle”, press the
/ button to highlight option, and then press the button
to select the option.• Press the
/ button to change the value; press the button to save the desired value.
• Press the
/ button and the button to select “Back/Exit” to return.
• Press the
button and / button to highlight “Clear Data” option, and then press the
button to select the option.• Entry “Clear Data”, press the
/ button to highlight “Sure” option, and then press the
button to select the option.• Press the
/ button and the button to select “Back/Exit” to return.
No.NameLCD DisplaySoundControl
1Power on
V0.00
0
0
% SpO
2
Artery
Check
BPM
288
M
ACTPPG
Beep for 2 secondsPress buttonSoftware version
start measure
2
Pulse search The icon ickersXX
3
Max./ Min SpO2 and Pulse rate warningShow or in red icon.Reects upper or lower alarm limits.Beep-beep sounded repeatedlyDefaultMute for temporary, will Beep-beep sounded after two minutes. modeMute before power o modeBar Chart• Entry “Data History”, press the / button to highlight “Bar Chart” option for viewing statistics of SpO2, and then press the button to select the option.• Entry “SpO2 Memory Data”, press the button to go back to the monitoring screen. Memory Transfer to ROSSMAX Software
• Entry into “Data History”, press the
/ button to highlight “Memory Transfer” option, and then press the button to start data transmission for a few seconds.Note: Before entering the data history mode, the icon should collect more than 1 data.
4
Probe failure alarm
ACTPPG
The icon shows on the screenBeep-beep sounded repeatedly
X
5
Probe connect failure alarm
ACT PPG
The icon shows on the screenBeep-beep sounded repeatedly
x
6
Low
battery alarm
ACT PPG
The icon shows on the screenBeep-beep sounded repeatedly for 1 minute and power o
x
7
Automatic
O
X
Beep-beep sounded and then power o After pulse is unde-tectable for around 1 minute.
The SA300 is intended for use in the electromagnetic environment specied below. The customer or the user of the SA300
should assure that is used in such and environment.
Immunity testIEC 60601 test levelCompliance levelElectromagnetic environment-guidance
Conducted RF
IEC 61000-4-6
3 Vrms:
0,15 MHz – 80 MHz
6 Vrms: in ISM and ama-
teur radio bands between
0,15 MHz and 80 MHz
80 % AM at 1 kHz
3 Vrms:
0,15 MHz – 80 MHz
6 Vrms: in ISM and ama-
teur radio bands between
0,15 MHz and 80 MHz
80 % AM at 1 kHz
Portable and mobile RF communications equip-
ment should be used no closer to any part of the
SA300 including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance:
d = 1,2 √P, d = 1,2 √P 80MHz to 800 MHz, d = 2,3 √P
800MHz to 2,7 GHz
IN0SA300000000031
Warranty CardThis instrument is covered by a 2 years guarantee from the date of purchase, batteries and accessories are not included. The guarantee is valid only on presentation of the guarantee card completed by the dealer confirming date of purchase or the receipt. Opening or altering the instrument invalidates the guarantee. The guarantee does not cover damage, accidents or non-compliance with the instruction manual. Please contact your local seller/dealer or www.rossmax.com.Customer Name: __________________________________________________________Address: _________________________________________________________________Telephone: _______________________________________________________________E-mail address: ____________________________________________________________Product Information: Date of purchase: __________________________________________________________Store where purchased: __________________________________________________________________________
Rossmax InnoTek Corp.
12F., No. 189, Kang Chien Rd., Taipei, 114, Taiwan.
CMC Medical Devices & Drugs S.L.
C/ Horacio Lengo N° 18, CP 29006, Málaga, Spain
Radiated RF IEC
61000-4-3
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m).
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2: These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reec-
tion from structures, objects and people.
More information on EMC compliance of the device may be obtained from Rossmax using the contacts shown in this manual.
The text is subject to change without further notice.
View the manual for the Rossmax SA300 here, for free. This manual comes under the category heart rate monitor and has been rated by 1 people with an average of a 9.5. This manual is available in the following languages: English. Do you have a question about the Rossmax SA300 or do you need help? Ask your question here
Need help?
Do you have a question about the Rossmax and is the answer not in the manual?
Receive updates on solutions
Get answers to your question
Question and answer
Number of questions: 0
The Rossmax SA300 is a heart rate monitor that provides accurate and reliable readings. It is designed to track your heart rate during physical activity or at rest, helping you monitor your cardiovascular health. The device features a clear display that shows real-time heart rate data, allowing you to easily track your fitness progress.
With its comfortable and adjustable strap, the Rossmax SA300 can be worn during various activities without causing discomfort. The monitor is compact and lightweight, making it convenient to carry around wherever you go. It also has a long-lasting battery life, ensuring that you can rely on it for extended use.
In addition, the Rossmax SA300 is user-friendly, with simple controls that make it easy to operate. Whether you are a fitness enthusiast or someone looking to monitor their heart health, this device is a practical and efficient option. Invest in the Rossmax SA300 to keep track of your heart rate and stay on top of your overall well-being.