View the manual for the Philips HeartStart FR2+ M3861A here, for free. This manual comes under the category defibrillators and has been rated by 1 people with an average of a 5.9. This manual is available in the following languages: English. Do you have a question about the Philips HeartStart FR2+ M3861A or do you need help? Ask your question here
The Philips HeartStart FR2+ M3861A is a defibrillator designed to deliver life-saving shocks to individuals experiencing sudden cardiac arrest. This compact and portable device is equipped with clear voice instructions and visual cues to guide users through the resuscitation process. It features SMART Biphasic technology, which adjusts the shock strength based on the patient's specific needs. The FR2+ is designed for ease of use in a variety of settings, including homes, schools, and medical facilities. It is equipped with pre-connected SMART Pads II, which simplifies the process of applying the electrodes to the patient's chest. Additionally, the device performs daily self-tests to ensure that it is ready for use at a moment's notice. Overall, the Philips HeartStart FR2+ M3861A is a reliable and user-friendly defibrillator that provides critical support in emergency situations. Its advanced technology and intuitive design make it an essential tool for anyone looking to effectively respond to sudden cardiac arrest.
| Brand | Philips |
| Model | HeartStart FR2+ M3861A |
| Product | defibrillator |
| Language | English |
| Filetype | User manual (PDF) |
Can't find the answer to your question in the manual? You may find the answer to your question in the FAQs about the Philips HeartStart FR2+ M3861A below.
To perform a self-test on the Philips HeartStart FR2+ M3861A defibrillator, follow these steps: 1. Ensure that neither the defibrillator pads nor the FR2+ ECG assessment module are connected to the device. 2. Insert the battery into the defibrillator. 3. The main menu will be displayed on the screen, and a two-part self-test will automatically run unless you make another selection from the menu within 10 seconds. 4. The self-test consists of an automatic part and an interactive part. 5. If the menu screen does not appear when the battery is inserted, it could be due to the defibrillator pads being attached to a patient, the FR2+ ECG assessment module being connected, or the battery being completely depleted. 6. To run the battery insertion self-test, remove and reinstall the battery. 7. The screen will indicate whether or not a data card is installed and display the remaining recording time on the card. 8. Use the lower Option button to navigate through the menu items and the upper Option button to select or scroll through the settings. 9. During the automatic part of the self-test, a bar will fill in on the screen to indicate the progress. 10. If the self-test detects a problem, a message will be displayed, and the Status Indicator will show a flashing or solid red X. 11. In such cases, replace the battery with a new one and repeat the self-test. If the second self-test also fails, contact Philips for technical support. Please note that these steps are provided as a general guide. For detailed instructions and specific troubleshooting scenarios, refer to the manual of the Philips HeartStart FR2+ M3861A defibrillator.
If the defibrillator pads do not adhere well to the patient's skin during use, follow these steps: 1. Ensure that the patient's skin is clean, dry, and free from any lotions, oils, or sweat. 2. Check the expiration date of the defibrillator pads. If they are expired, replace them with a new pair. 3. Verify that the defibrillator pads are properly positioned on the patient's chest according to the instructions provided in the manual. 4. Press firmly on the pads to ensure good contact with the skin. Apply sufficient pressure to ensure adhesion without causing discomfort to the patient. 5. If the pads still do not adhere well, try removing any hair or excessive moisture from the patient's chest using a dry cloth or towel. 6. Consider using an alternative pad placement location if the current location is not providing adequate adhesion. 7. If none of the above steps resolve the issue, replace the defibrillator pads with a new pair from a different package. 8. If the problem persists, contact Philips customer support for further assistance. Remember, ensuring proper adhesion of the defibrillator pads is crucial for effective delivery of the electrical shock during defibrillation.
To change the default configuration settings on the Philips HeartStart FR2+ M3861A defibrillator, follow these steps: 1. Access the configuration menu by pressing the Setup button on the defibrillator. 2. Navigate through the menu options using the arrow buttons until you find the "Configuration" or "Setup" option. 3. Select the "Configuration" or "Setup" option by pressing the Enter or Select button. 4. Within the configuration menu, you will find a list of parameters that can be modified. 5. Use the arrow buttons to scroll through the parameters and select the one you want to change. 6. Once you have selected a parameter, use the arrow buttons to adjust the settings according to your preference. 7. Some parameters may have multiple options, such as speaker volume or ECG display. 8. After making the desired changes, confirm the selection by pressing the Enter or Select button. 9. Repeat steps 5-8 for any other parameters you wish to modify. 10. Exit the configuration menu by pressing the Setup button again or navigating to the "Exit" or "Done" option. Please note that the specific configuration settings available may vary depending on the software version of your FR2+ defibrillator. For detailed instructions and specific parameter options, refer to the manual of the Philips HeartStart FR2+ M3861A defibrillator.
When using the Philips HeartStart FR2+ M3861A defibrillator on an infant or child in cardiac arrest, follow these steps: 1. Press the On/Off button to turn on the defibrillator and follow the voice and text prompts provided by the FR2+. 2. Remove clothing from the patient's upper body and, if necessary, wipe any moisture from the patient's skin. Clip or shave excessive chest hair if needed. 3. If the patient appears to be under eight years of age or weighs 55 lb. (25 Kg) or less, use the M3870A FR2 infant/child reduced-energy defibrillator pads if available. If these pads are not available or if the child appears older or larger, use adult defibrillator pads. Do not delay treatment to determine the child's exact age or weight. 4. Open the defibrillator pads package and remove the protective backing from the pads. 5. Place the sticky side of each pad on the patient's bare skin as shown in the drawing on each pad. 6. Insert the defibrillator pads connector firmly into the defibrillator's connector socket, indicated by a flashing light at the top left of the FR2+. 7. Follow your local protocol for providing infant/child CPR while a bystander calls EMS and brings the FR2+. If no bystander is available, provide 1-2 minutes of CPR before calling EMS and retrieving the FR2+. 8. If you witnessed the child's collapse, call EMS immediately and then get the FR2+. Please note that these steps are a general guide. For detailed instructions and specific guidelines, refer to the manual of the Philips HeartStart FR2+ M3861A defibrillator.
When using the Philips HeartStart FR2+ defibrillator in an emergency situation, here are some general troubleshooting tips: 1. Self-Test: Before using the defibrillator, ensure that it has passed the self-test. The self-test consists of an automatic part and an interactive part. If any part of the self-test indicates a problem, do not use the defibrillator and contact Philips for technical support. 2. Status Indicator: Pay attention to the status indicator on the defibrillator. If you see a flashing or solid red X on the status indicator, it may indicate a low battery, storage outside the recommended temperature range, or an error detected during the self-test. Take appropriate action based on the specific situation, such as replacing the battery or contacting Philips for assistance. 3. Audio Signals: Listen for audio signals from the defibrillator. If you hear chirping sounds, it may indicate a low battery or a self-test failure. Take the necessary steps, such as replacing the battery or performing a battery insertion self-test, as instructed in the manual. 4. Text Prompts: Pay attention to any text prompts displayed on the defibrillator's screen. These prompts may provide important information about the readiness of the defibrillator or any self-test failures. Follow the instructions provided in the prompts and take appropriate action accordingly. Remember, if you encounter any issues or have doubts about the functionality of the defibrillator, it is crucial to seek professional assistance and not use the device until the problem is resolved.
To turn on the Philips HeartStart FR2+ M3861A defibrillator, follow these steps: 1. Press the On/Off button located on the defibrillator. 2. Follow the instructions provided by the FR2+ voice and text prompts. 3. Prepare the patient by removing clothing from their upper body. If necessary, wipe any moisture from the patient's skin and clip or shave excessive chest hair. 4. If the patient is under eight years of age or weighs less than 55 lb. (25 Kg), use M3870A FR2 infant/child reduced-energy defibrillator pads if available. If these pads are not available or if the child appears older or larger, use adult defibrillator pads. Do not delay treatment to determine the child's exact age or weight. 5. Open the defibrillator pads package and remove the protective backing from the pads. 6. Place the sticky side of each pad on the patient's bare skin as shown in the drawing on each pad. 7. Insert the defibrillator pads connector firmly into the defibrillator's connector socket, indicated by a flashing light at the top left of the FR2+. Please note that these steps are based on the information provided in the manual for the Philips HeartStart FR2+ M3861A defibrillator. Always refer to the manual for detailed instructions and guidelines.
If the status indicator on the Philips HeartStart FR2+ M3861A defibrillator is not working, you can follow these steps to troubleshoot the issue: 1. Check the battery: Ensure that the battery is properly inserted and charged. If the battery is missing or completely depleted, install a new M3863A FR2 standard battery or a charged M3848A FR2+ rechargeable battery. 2. Perform a battery insertion test: Remove and reinstall the battery, then perform the battery insertion self-test. If the test fails, try installing a new M3863A FR2 standard battery or a charged M3848A FR2+ rechargeable battery and repeat the test. If the test fails again, do not use the FR2+. 3. Check for visible damage: Inspect the defibrillator for any visible damage. If it appears to be damaged, do not use it. 4. Contact technical support: If you have tried the above steps and the status indicator is still not working, contact Philips for technical support. Please note that these troubleshooting steps are based on the information provided in the manual for the Philips HeartStart FR2+ M3861A defibrillator. Always refer to the manual for detailed instructions and guidelines.
To attach the disposable defibrillator pads to the Philips HeartStart FR2+ M3861A defibrillator, follow these steps: 1. Open the package containing the defibrillator pads. 2. Remove the protective backing from the pads to expose the adhesive side. 3. Position the pads on the patient's bare chest, following the placement diagram provided on the pads. 4. Ensure that the pads are firmly attached to the patient's skin, making sure they are centered and aligned correctly. 5. Connect the pads to the defibrillator by inserting the pads' connector firmly into the defibrillator's connector socket. 6. Verify that the pads are securely connected by checking for a flashing light at the top left of the FR2+. 7. Once the pads are attached, follow the voice and text prompts provided by the defibrillator to continue with the appropriate actions. Please note that these steps are based on the information provided in the manual for the Philips HeartStart FR2+ M3861A defibrillator. Always refer to the manual for detailed instructions and guidelines.
If the text prompts on the screen of the Philips HeartStart FR2+ M3861A defibrillator are not displaying correctly, you can try the following steps to troubleshoot the issue: 1. Ensure proper positioning: Make sure that the defibrillator pads are correctly positioned on the patient's chest according to the diagram on the back of the pads. 2. Check the connector: Verify that the defibrillator pads connector is fully inserted into the connector socket on the defibrillator. 3. Press the pads firmly: Press the defibrillator pads firmly onto the patient's chest to ensure proper contact. 4. Replace the pads if necessary: If the text prompts continue to display incorrectly, consider replacing the defibrillator pads with new ones. 5. Contact technical support: If the issue persists after trying the above steps, it is recommended to contact Philips for technical support. Please note that these troubleshooting steps are based on the information provided in the manual for the Philips HeartStart FR2+ M3861A defibrillator. Always refer to the manual for detailed instructions and guidelines.
The M3861A model of the Philips HeartStart FR2+ defibrillator does not have the option for manual charge in advanced mode due to its design and intended use. Here's why: 1. Simplified operation: The M3861A model is designed to provide simplified operation and ease of use, especially in emergency situations. It focuses on providing clear and concise text prompts to guide the user through the resuscitation process. 2. Automated energy selection: The defibrillator automatically selects the appropriate energy level for each shock based on the patient's needs. This eliminates the need for manual charge adjustment in advanced mode, as the device is programmed to deliver the necessary energy for effective defibrillation. 3. Safety considerations: The absence of manual charge in advanced mode helps to ensure the safety of both the user and the patient. It reduces the risk of accidental or inappropriate energy delivery, as the device is designed to deliver the appropriate energy level automatically. 4. User-friendly interface: The M3861A model focuses on providing a user-friendly interface with clear text prompts, making it easier for users with varying levels of training and experience to operate the defibrillator effectively. Please note that these reasons are based on the information provided in the manual for the Philips HeartStart FR2+ M3861A defibrillator. Always refer to the manual for detailed instructions and guidelines.
Yes, the manual of the Philips HeartStart FR2+ M3861A is available in English .
Is your question not listed? Ask your question here